Lift Assist for Wheelchairs User Manual
Everything in the printed manual, as one page: safety, setup, placing it on the wheelchair, the remote control, sitting down and standing up, care, troubleshooting, specifications and warranty.
Download PDF manual8.1 MB
Welcome to your Lift Assist for Wheelchairs.This page is the complete user manual. Read the Safety chapter before first use, then follow Using the Lift Assist Wheelchair: setup, placing it on the wheelchair, the remote control, sitting down and standing up. Use the chapter bar above to jump, or search for any word.
Chapter 1Introduction
About this manual
This manual provides the information necessary to operate the Lift Assist Wheelchair in a safe and efficient manner. Please review and understand this manual before operating the system. If any part of this manual is unclear, contact your supplier or SitnStand Customer Support for clarification.
Life Assistant SitnStand for Wheelchairs product overview
Lift Assist Wheelchair – Portable Smart Rising Seat is a lightweight, portable inflatable system for wheelchairs that is safe, simple and easy to operate. It comprises an electronic control system, multiple air valves, multiple air sealed layers, and a portable air pump that supplies the required pressure and air capacity to assist standing up and sitting down with minimum or no help required.

Contents
- 1 Lift Assist Wheelchair unit
- 1 visco / memory foam cushion (2 inch / 5 cm), embedded in the sitting area
- 1 Li-ion 14.4 VDC rechargeable battery
- 1 wall battery charger with 1 m cable and connector
- 1 user manual
Warnings, cautions and notes
Three types of special messages appear in this user manual:
WARNING indicates the possibility of injury to the patient or operator.
CAUTION indicates a condition that may lead to equipment malfunction.
NOTE provides other important information.
Conditions for use
Intended use
The Lift Assist Wheelchair is a powered inflatable cushion intended to assist patients suffering from mobility difficulties to rise from a seated position to a standing position, as well as to sit down from a standing position.
The Lift Assist Wheelchair is intended to be used in the homecare environment while performing daily activities, as well as in medical care facilities.
Intended users
The Lift Assist Wheelchair is intended for the home care environment, to be used by elderly people and/or people with compromised mobility or difficulties that require the use of a wheelchair.
The Lift Assist Wheelchair is intended to be used by people with chronic or acute mobility difficulty who use a wheelchair.
The Lift Assist Wheelchair is intended to be used by cognitive impairment patients who sit on a wheelchair while assisted by a caretaker. They should familiarize themselves with the instructions in this manual.
Contraindications
- Do not use the Lift Assist Wheelchair while under the influence of narcotics or alcohol.
- Do not use the Lift Assist Wheelchair while holding a child or infant.
- Do not use the Lift Assist Wheelchair inside a car.
- Do not use the Lift Assist Wheelchair when undressed.
- In the case of cognitive impairment, use the Lift Assist Wheelchair only with a caregiver.
Chapter 2Safety
General warnings and cautions
WARNING: DO NOT USE PRIOR TO READING AND UNDERSTANDING THIS MANUAL.
WARNING: To avoid the potential of electrical shock, use only the designated charger provided with the device. Using any other charger may cause an electrical and/or electromagnetic hazard.
WARNING: Do not move the wheelchair until the cushion is fully deflated.
WARNING: To avoid the potential of electric shock, the power supply must be disconnected from the power mains prior to cleaning.
WARNING: To avoid the potential of electric shock, contact your supplier or SitnStand when a malfunction occurs.
WARNING: Do not inflate under the bottom of a table or other item that interferes with the cushion inflation.
NOTE: The only user-serviceable component in the system is the battery.
CAUTION: Changes or modifications not expressly approved by SitnStand can affect the safety and effectiveness of the system, and will invalidate the system's warranty.
CAUTION: Do not operate with damaged cords or plugs. If damaged, please contact the company.
CAUTION: At the end of its useful life, the system and battery must be disposed of in accordance with local law and/or code concerning electrical and electronic equipment.
Chapter 3Labels and symbols
Chapter 4Using the Lift Assist Wheelchair
4.1 Setup instructions
Setting up the Lift Assist Wheelchair is performed by the user. The setting up process consists of three steps:
- Charging the battery for the first time.
- Placing the Lift Assist Wheelchair on a wheelchair.
- Connecting the battery.
A. Charging the battery for the first time
Charge the battery for at least 12 hours before use:
- Remove the battery from the box.
- Plug the battery charger into an electric wall socket.
- Connect the battery charger connector to the battery charging socket.
- When the green light is on, the battery is fully charged.
- When the red light is on, the battery is not fully charged.
NOTE: To prevent falling over any tangled cable, place the battery close to the charger.

B. Placing the Lift Assist Wheelchair on a suitable seat
Before you start: make sure that the wheelchair brakes are locked.
- Place the Lift Assist Wheelchair folded on the wheelchair (Figure 3).
- Insert the back part (the air supply unit) into the space between the back and the seat of the wheelchair (Figure 4).


- Place the air supply unit's straps on the wheelchair's handles and tighten them (Figure 5).
- When not in use, we recommend attaching the remote control to a velcro strip on the wheelchair's side handles (Figures 6 and 7).



C. Inserting the battery in its place
Slide the charged battery into its place within the air supply unit, as illustrated in Figure 8.
The Lift Assist Wheelchair is now in standby mode, and is ready for use.

4.2 Controller indicator modes
4.3 Sitting down
Before sitting down: make sure that the wheelchair brakes are locked.
- Inflate the cushion using the controller to 1, 2 or 3 layers (whichever is most comfortable for you).
- Sit comfortably in the center.
- When you feel steady, start deflating the cushion by pressing the Down Arrow button on the controller.
- You can now sit down comfortably.
4.4 Standing up
Before standing up: make sure that the wheelchair brakes are locked.
- Place your feet on the floor, close to the chair.
- Sit upright.
- Spread your feet.
- Inflate the cushion by pressing the Up button.
- Now you can easily stand up.
NOTE: In case of cognitive impairment, the Lift Assist Wheelchair should only be operated by a caregiver.
The Lift Assist Wheelchair has 3 layers; however, inflation can be stopped at any time when the level of elevation is sufficient.
Important: Do not attempt to stand up before the layer is totally inflated and there is no sound from the air pump.
The Lift Assist Wheelchair may remain inflated until the next time it is used to sit down.
4.5 Replacing the battery
When the Charge Battery indicator turns orange, the battery should be charged.
Open the zipper covering the air supply unit and remove the battery by sliding it towards you, as illustrated in Figure 10.
The battery is connected by a magnet to the air supply unit.
NOTE: Don't remove the battery while the Lift Assist Wheelchair is in Working Mode (see controller indicator modes).

4.6 Charging the battery
- Plug the battery charger into an electric wall socket.
- Connect the battery charger connector to the battery charging socket.
- When the red light on the charger changes to green, the battery is fully charged.
- When the red light is on, the battery is not fully charged.

Chapter 5Maintenance
In order to maintain the Lift Assist Wheelchair system, please note the following:
- Keep the Lift Assist Wheelchair away from dust or dirt.
- Use a soft dry brush to clean the fabric surfaces (e.g. seat area).
- Do not iron the textile.
- Use a damp cloth to clean the plastic surfaces (e.g. air supply unit).
- Do not use solvents or detergents to clean any part.
- Do not allow liquids near the system openings.
- Keep dry.
WARNING: To avoid potential of electrical shock, the power supply must be disconnected from the power mains prior to cleaning.
WARNING: To avoid the potential of electric shock, contact your supplier or Life Assistant Ltd (SitnStand) when a malfunction occurs.
NOTE: The only user-serviceable component in the system is the battery.
NOTE: No portion of the system should be immersed in water or other fluids.
NOTE: Never spray or otherwise apply any detergent fluid directly into the openings or seams of the system. Always apply the detergent fluid to the cloth instead.
Chapter 6Troubleshooting
Before addressing the troubleshooting table, please check and confirm the following:
- The battery is attached to the Lift Assist Wheelchair.
- The battery is charged (low battery indicator is off, standby/ready indicator is on).
Basic troubleshooting
When addressing a malfunction, as an initial action please follow Step 1 described in the table below. In case of an unsolved malfunction, follow Step 2.
For more information
Life Assistant LTD (SitnStand)
6 HaParsa St. Afula, 1829306 ISRAEL
email: info@sitnstand.com
web: www.sitnstand.com
NOTE: If you experience any unusual symptoms, discomfort, or difficulties while using the device, stop using it and consult a healthcare professional.
Any serious incident (i.e. major injury) that has occurred in relation to the device should be reported to the manufacturer and to the competent authority of the Member State in which the user is established.
Chapter 7Operating specifications
NOTE: Unless otherwise indicated, all specifications are subject to change without notice. Specifications and test methods will be made available upon request.
7.1 Environmental
Transport and storage environment
Operation environmental conditions
7.2 Electrical
CAUTION: At the end of its useful life, the system and battery must be disposed of in accordance with local law and/or code concerning electrical and electronic equipment. Do not discard to standard trash bin.
7.3 Physical and operational
Green LED: system ready
Orange LED: low battery
Red LED: error
Chapter 8Limited warranty
Life Assistant LTD (SitnStand) warrants this product, including the battery and the textile/fabric, for a period of two years from the date of purchase.
The warranty covers defects in materials and workmanship, comprising the system, assemblies, components and labor.
The warranty is not valid in the case of abnormal wear and tear, willful damage, neglect, misuse or unauthorized repair or alteration of the product, or any of its parts or accessories.
The following statements do not affect the consumer's statutory rights. This product is sold on the understanding that, in the event of any defect of manufacture or material appearing within two years of the date of receipt of this product, the defect will be corrected free of charge on condition that evidence of date of ownership is provided. Such evidence may be an invoice, delivery note, etc. and evidence that the defect is not attributable to accidental damage either in transit or otherwise, misuse or unauthorized repair or alteration.
Your product record
Keep these details for any warranty claim:
Chapter 9EMC statement
Electromagnetic compatibility declarations, as required for medical electrical equipment. Mainly relevant for institutions, installers and regulators.
Warnings
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided as it may result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
WARNING: Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment may result in increased electromagnetic emissions or decreased electromagnetic immunity of this equipment, resulting in improper operation.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Lift Assist Wheelchair, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment may result.
Electromagnetic emissions
Lift Assist Wheelchair is intended for use in the electromagnetic environment specified in the following table. This is not a life-sustaining device.
The user and/or installer of the unit must ensure that it is used in such an environment.
| Emissions test | Compliance | Electromagnetic environment – guidance |
|---|---|---|
| RF emissions CISPR 11 | Group 1, Class B | The Lift Assist Wheelchair uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. |
| Harmonic emissions IEC 61000-3-2 | N/A | The Lift Assist Wheelchair is suitable for use in all establishments, including domestic establishments and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes. |
| Voltage fluctuations and flicker IEC 61000-3-3:2013 | Complies |
The remote control
Orange light. On when the battery needs recharging; when the green light is also off, the battery is depleted (see the modes below).
Green light. Flashes slowly when the device is ready for use, and quickly while it is operating.
Red light. Flashes quickly together with the green light when the device has identified a system fault. See the Troubleshooting chapter.
Press to inflate the cushion one layer at a time. You can stop at any layer.
Press to deflate the cushion, for example once you are seated and feel steady.

Reset: pressing both buttons together for 10 seconds resets the system. Use it only as described in the Troubleshooting chapter.